endogirl

 

ENDD13-0910: The Importance of Pulse Oximetry

Author: Sharon Lesser, RN
1.0 contact hours

How One Technological Advance Changed Medicine

Pulse oximetry is a simple, non-invasive method of monitoring the percentage of hemoglobin (Hb) that is saturated with oxygen. The pulse oximeter consists of a probe attached to the patient’s finger or earlobe, which is linked to a computerized unit. The unit displays the percentage of Hb saturated with oxygen together with an audible signal for each pulse beat, a calculated heart rate in some models, and a graphical display of blood flow past the probe. Audible alarms that can be programmed by the user become clinically cyanosed.

Since the device is non-invasive and allows immediate and real time monitoring, it use has expanded to include other purposes such as screening, diagnosis, patient follow-up, and self-monitoring.

History

In 1935, Matthes developed the first 2-wavelength ear O2 saturation meter with red and green filters, later switched to red and infrared filters. This was the first device to measure O2 saturation.

In 1949, Wood added a pressure capsule to squeeze blood out of the ear to obtain zero setting in an effort to obtain absolute saturation value when blood was remitted. This concept is similar to today’s conventional pulse oximetry but suffered due to unstable photocells and light sources.

This method was not used clinically. In 1964, Shaw assembled the first absolute reading ear oximeter by using eight wavelengths of light. Commercialized by Hewlett Packard, it use was limited to pulmonary functions and sleep laboratories due to the cost and the size.

Pulse oximetry was developed in 1972, by Aoyagi at Nihon Kohden using the ratio of red to infrared light absorption of pulsating components at the measuring site. It was commercialized by BIOX/Ohmeda in 1981 and Nellcor in 1983. Nellcor incorporated in 1982, and introduced it into the US operating room market in 1983. Prior to its introduction, a patient’s oxygenation was determined by a painful arterial blood gas, a single point of measure which typically took a minimum of 20-30 minutes of processing by the lab. (Remember that in the absence of oxygenation, damage to the brain starts in five minutes, with brain death in another 10-15 minutes.)

In the U.S. alone, approximately $2 billion was spent annually on this measurement. With the introduction of pulse oximetry, a non-invasive, continuous measure of patients' oxygenation was possible, revolutionizing the practice of anesthesia and greatly improving patient safety. Prior to its introduction, studies in anesthesia journals estimated U.S. patient mortality as a consequence of undetected hypoxemia at 2,000 to 10,000 deaths per year, with no known estimate of patient morbidity.

By 1987, the standard of care for the administration of a general anesthetic in the U.S. included pulse oximetry. From the operating room, the use of the pulse oximetry rapidly spread throughout the hospital, first in the recovery room, and then into the various intensive care units.

How Does a Pulse Oximeter Work

A pulse oximeter consists of a peripheral probe together with a microprocessor unit, displaying a waveform — the oxygen saturation and the pulse rate.

Most oximeters have an audible pulse tone, the pitch of which is proportional to the oxygen saturation — which can be useful when you cannot see the display. The probe is placed on a peripheral part of the body such as a finger, toe, earlobe or the nose.

Within the probe are two light-emitting diodes (LEDs). One is visible red spectrum and the other is infrared spectrum. The beams of light pass through the tissues; some light is absorbed by the blood and soft tissue depending on the concentration of hemoglobin.

Saturation values are averaged out over five to 20 seconds. The pulse rate is also calculated from the number of LED cycles between successive pulsatile signals and average out over a similar variable period of time, depending on the particular monitor.

Calibration and Performance

Oximeters are calibrated during manufacture and automatically check their internal circuits when they are turned on. They are accurate in the range of oxygen saturations of 70 percent-100 percent (+/- 2 percent), but less accurate under 70 percent. The pitch of the audible pulse signal falls with reducing values of saturation.

The size of the pulse wave is displayed graphically. Some models automatically increase the gain of the display when the flow decreases; in these, the display may prove misleading. The alarms usually respond to a slow or fast pulse rate or an oxygen saturation below 90 percent. At this level, there is a marked fall in PaO2, representing serious hypoxia.

Limitations/Pitfalls

Pulse oximetry is a measure solely of oxygenation, not of ventilation, and is not a substitute for blood gases checked in the laboratory. Pulse oximetry does not give an indication of carbon dioxide levels, blood pH, or sodium bicarbonate levels.

False low readings may be caused by hypoperfusion of the extremity being used for monitoring (often because the part is cold or because of the vasoconstriction secondary to the use of vasopressor agents); incorrect sensor application; highly calloused skin; and movement (such as shivering), especially during hypoperfusion. Ambient light, abnormal hemoglobin; pulse rate and rhythm and cardiac function can also cause abnormal pulse oximeter reading. To ensure accuracy, the sensor should return a steady pulse and/or waveform. Falsely high or falsely low readings will occur when the hemoglobin is bound to something other than oxygen. In cases of carbon monoxide poisoning, the falsely high reading may delay the recognition of hypoxemia. Cyanide poisoning can also give a false high reading.

What factors cause errors in the pulse oximeter?

Abnormal hemoglobin -- Blood may contain abnormal hemoglobin such as carboxyhemogolobins and methemoglobin, which do not contribute to oxygen delivery.

Medical dyes -- If dyes such as cardio green, intravascular dyes and indocyanine green have been injected into the blood, they may influence the level of transmission of the red and infrared light.

Manicure and pedicure -- If the users wear nail polish, it may absorb the light emitted from the LED and change the light transmitted through the body, influencing the values calculated.

Major body motion -- Body motion may cause noise that affects the values calculated. When noise -- including that caused by body motion -- is reduced, the reliability of the values calculated falls, and the pulse oximeter displays a warning.

Blood flow blocked by pressure on the arms or fingers -- The pulse oximeter measures oxygen saturation based on changes in the blood flow. Therefore, if the blood is blocked, correct measurement becomes impossible. In addition, if the fingers are flexed at a uniform place, the pulse oximeter may interpret the pressure changes in pulse rate, causing errors.

Peripheral circulatory failures -- The pulse oximeter utilizes blood flow to monitor changes in the amount of light transmitted to calculated values. If peripheral blood flow is reduced, adequate data may not be obtained, and the result is inaccurate measurement. In this case, it is necessary to promote blood flow by massaging or warming the fingers.

Excessive ambient light -- The pulse oximeter can usually cancel out the effects of ambient light. However, if the ambient light is too strong, the device will not be able to cancel out the effects and this may cause errors.

Ambient electromagnetic waves -- If electric appliances such as television, mobile telephones, or medical devices that produce high levels of electromagnetic waves are used near the pulse oximeter, the electromagnetic waves from these devices may interfere with accurate measurement.

Probe attached incorrectly -- If the probe is not attached properly, it may detect a variety of noise, resulting in inaccurate measurement.

What exactly does the pulse oximeter measure?

Blood carries oxygen in two forms. The majority is bound to hemoglobin (oxyhemoglobin), and the rest is dissolved in the aqueous phase of blood (the plasma). The pulse oximeter measures the saturation of hemoglobin with oxygen. This is expressed as a percent saturation. Each gram of normal hemoglobin can hold 1.34 milliliters of oxygen. The dissolved fraction is dependent upon the partial pressure of oxygen. As the partial pressure increases, the dissolved fraction of oxygen increases. For each 1 mm/Hg pressure of oxygen partial pressure, 0.003 milliliters dissolves in the plasma. So under normal conditions, each 100 mL of blood contains about 20 mL of oxygen bound to hemoglobin and about 0.3 mL dissolved in plasma. The dissolved fraction is available to tissues first, and then the fraction bound to hemoglobin. So as tissues metabolize oxygen, or if oxygen becomes difficult to pink up through the lungs, the dissolved oxygen and the hemoglobin-bound oxygen will eventually become depleted.

The dissolved oxygen can be measured by arterial blood gas analysis, but this is not yet practical for field application. This fraction is not measured by pulse oximetry. The pulse oximeter waits to sense the pulse of capillary blood from the side of the capillaries, then, using two different wavelengths of light, calculates the percent of oxyhemoglobin from the total hemoglobin present. If oxygen transfer across the lungs or lung function is compromised and tissues continue to metabolize oxygen, the percentage of oxyhemoglobin will decrease. This becomes our quantitative indicator of hypoxia.

When should supplemental oxygen be administered despite a normal pulse ox? Oxygen should be delivered whenever the underlying problem is suspected to be organ or tissue ischemia. This would include things like stroke, intracerebral bleed, head injury, altered mental status from any cause, chest pain of suspected cardiac origin, cardiac dysrhythmias, shortness of breath from any cause, vascular emergencies like aortic dissection or aneurysm or vascular occlusions, shock from any cause, sickle cell disease and multisystem trauma.

What respiratory function will pulse ox not measure? The pulse oximeter measures oxygenation. It does not measure ventilation. Ventilation is the process of removing carbon dioxide from the blood. Hypoventilation from any cause will result in the accumulation of carbon dioxide in the blood. This leads to respiratory acidosis. If hypoventilation goes on long enough, blood oxygen will begin to deplete and the pulse oximeter oxygen saturation will begin to decrease. However, short of apnea, the rate of carbon dioxide accumulation and the development of respiratory acidosis may be greater then the rate of onset of hypoxia by pulse oximetry. The assessment of adequate ventilation is based on respiratory rate and depth.

Should I hold on administering oxygen in order to check a baseline pulse ox? The decision to administer supplemental oxygen is based on available history and initial examination. If your first impression is that the patient is really sick, then they probably are. So don’t feel compelled to document a baseline pulse ox in a patient you think is in trouble. If the flow of the call results in an off-oxygen pulse ox reading being available, it certainly can be useful to correlate to symptom severity, but don’t hold off on oxygen waiting to get this value.

Where this might be useful is if you have a patient who is symptomatic and looks OK for the moment, but you would like to get a better assessment of the severity of the symptoms.

Practical Tips for Successful Use of Pulse Oximetry

Oximetry is not a complete measure of respiratory sufficiency. A patient suffering from hypoventilation (poor gas exchange in the lungs) given 100 percent oxygen can have excellent blood oxygen levels while still suffering from respiratory acidosis due to excess carbon dioxide.

It is also not a complete measure of circulatory sufficiency. If there is insufficient blood flow or insufficient hemoglobin (anemia), tissues can suffer hypoxia despite high oxygen saturation in the blood that does arrive.

It is important to remember that pulse oximetry is only one way of monitoring breathing. It is also necessary, as a minimum, to record respiratory rate, and if pulse oximetry is used, the amount of oxygen the patient is receiving must be recorded. As with all clinical assessments, you must look at the "whole picture."

Sharon Lesser, RN, is a pulmonary clinical nurse II in the department of pulmonary and critical care medicine at the University of Maryland Hospital in Baltimore, Md.

References

1. Grap MJ. Pulse oximetry. In: AACN Protocols for Practice: Technology Series. Aliso Viejo, Ca: American Association of Critical-Care Nurses; 1996.

2. Grap MJ. Pulse oximetry. Crit Care Nurse. 1998; 18:94-99.

3. Rutherford KA. Principles and application of pulse oximetry. Crit Care Nurs Clin Am. 1989; 1:649-657.

4. Robertson RE, Kaplan RE Another site for the pulse oximeter probe. Anesthesiology. 1991; 74:198.

5. Siegel MN, Garvenstein N. Preventing ambient light from affecting pulse oximetry. Anesthesiology. 1987; 67:280.

6. Zablocki AD, Rasch DK. A simple method to prevent interference with pulse oximetry by infrared heating lamps. Anesth Analg. 1987; 66:915.

7. Hanowell L, Eisele JH, Downs D. Ambient light affects pulse oximeters. Anesthesiology. 1987; 67:864-865.

8. Szarlarski NL, Cohen NH. Use of pulse oximetry in critically ill adults. Heart Lung. 1989; 18:444-453.

Carroll P. Using pulse oximetry in the home. Home Healthc Nurse. 1997; 15:88-97.

9. Tittle M, Flynn MB. Correlation of pulse oximetry and co-oximetry. Dimens Crit Care Nurs. 1997; 16:88-95.

10. Wahr JA, Tremper KK, Dlab M. Pulse oximetry. Respir Care Clin N Am. 1995; 1: 77-105.

11. New W. Pulse Oximetry. Journal Clinical Monitoring. 1985 April; 1 (2): 126-129.

12. Kelleher JF . Pulse Oximetry. Journal Clinical Monitoring 1989; 5: 37-62.

13. Tremper KK, Barker SJ. Pulse Oximetry. Anesthesiology 1989; 70: 98-108.

Take The Test! Click Here

homeadvisory boardauthorsonline coursescontact usMy Course Trackerresources
EndoNurse

All material on this site Copyright © 2012 Virgo Publishing, LLC. All rights reserved.
Please read our legal page before using this site. Privacy statement.